Calibration Columns
BULLETIN CC-08
Calibration Columns
Purpose
Neptune Calibration Columns provide a
fast, easy and economical means of checking
the flow rate of your chemical metering pump.
The most accurate test of metering pump flow rate
is to measure the drawdown rate on the suction side
while leaving the discharge undisturbed in its normal
steady state operating condition. Pump flow rate
verification on a periodic basis or after maintenance
is important to system accuracy.
The Calibration Column can also be used to determine
if check valves are worn or dirty.
It is recommended to size the Calibration Column for a
minimum of 30 seconds with a scale that will produce
an easily readable differential within the test period.
Features
Rugged PVC construction with slip on top
caps for top filling and easy cleaning
Shielded glass models available for acids
and strong chemicals
Calibration scales are protected from harsh
chemical by Mylar lamination
Specifications
FOR USE SCALE DIMENSIONS
MODEL
CAPACITY WITH PUMPS GRADUATION
NUMBER
UP TO INCREMENTS ABC
CC 100 100 ml. 3 gph 1 ml. 14" 1-5/8" 1/2" FNPT
CC 250 250 ml. 7 gph 2 ml. 15" 2-1/4" 1/2" FNPT
CC 500 500 ml. 16 gph 5 ml. 15-5/8" 2-3/4" 3/4" FNPT
CC 1000 1000 ml. 30 gph 5 ml. 26-3/8" 2-3/4" 3/4" FNPT
CC 4000 4000 ml. 120 gph 20 ml. 31-1/2" 5" 2" FNPT
B
A
C (INLET
& OUTLET)
Vista de pagina 0
1 2

Indice de contenidos

Pagina 1 - Calibration Columns

Calibration ColumnsBULLETIN CC-08Calibration ColumnsPurposeNeptune Calibration Columns provide afast, easy and economical means of checkingthe flow ra

Pagina 2 - Operating Instructions

Installation InstructionsThe recommended installation is shown below:1. Install the Calibration Column in the suction lineof the Chemical Feed Pump. T